          44:03:01:08.05.  Periodic measurement of medical fluoroscopic spot film devices. Periodic measurement of the entrance radiation exposure rate shall be performed by a qualified expert for both typical and maximum values. Such measurements shall be made triennially or after any maintenance of the system which might affect the radiation exposure rate. Results of these measurements shall be posted where any operator may have ready access to such results while using the fluoroscope. Results of the measurements shall include the roentgen per minute, as well as the technique factors used to determine such results. The name of the person performing the measurements and the date the measurements were performed shall be included in the results.

          Source: 26 SDR 96, effective January 23, 2000.

          General Authority: SDCL 34-21-4.1, 34-21-15.